Does a failure in a zero solenoid valve impact measurement accuracy of the analyser?

In a Continuous Emission Monitoring System (CEMS), even the smallest mechanical failure can mislead data and compliance efforts. An issue that often goes unnoticed but has significant impact is failure in the zero solenoid valve (SOV).
The role of the zero SOV is critical: during calibration or zero-checks, it ensures that only clean, zero air flows to the analyser.

But what happens when this valve leaks?
The failure causes zero air to mix continuously with the flue gas sample.

The result?
The analyser records emissions lower than they actually were. Because the analyser can’t distinguish the dilution, it measures a falsely “cleaner” reading, posing a serious compliance risk. Over time, this kind of dilution can not only distort emission reporting but also affect decisions that depend on reliable analyser output.

By being digitally connected, the system can detect inconsistent zero readings and unusual measurement patterns. This anomaly proves as a clear signal that something is wrong beyond the analyser’s core functions.

Forbes Marshall Digital’s EverSense for CEMS’s real-time data and pattern recognition forms a base for our specialists to identify the root cause. Once identified as a leakage in the zero SOV, a simple valve replacement restores measurement accuracy. The advantage of a connected, real-time monitoring system like EverSense, along with the complete Forbes Marshall Digital package is this: we don’t just help identify problems – EverSense along with our specialists help diagnose  problems correctly, which would have otherwise remained masked, leading to prolonged non-compliance and incorrect emissions data.

Click here to read more about the service.